HIP:YSI-600R:minicom
Jump to navigation
Jump to search
Setting-up minicom
The following minicom screen shows the correct configuration to use with a YSI sonde on comm 2 of the TS-7260 board.
------------------------------------------------------------------------ | A - Serial Device : /dev/ttyAM1 | | B - Lockfile Location : /var/lock | | C - Callin Program : | | D - Callout Program : | | E - Bps/Par/Bits : 9600 8N1 | | F - Hardware Flow Control : No | | G - Software Flow Control : No | | | | Change which setting? | ------------------------------------------------------------------------
Using minicom with Sonde
- Enabling VT100 emulation in the sonde will make it clearer to use the menu. To do this, set Main > Advanced > Advanced setup > VT100 emulation.
- Many of the sonde commands and menus available via RS-232 are documented in the YSI sonde manual. However, there are some that are worth noting here.
- Typing "run" at the sonde prompt will cause the sonde to print data lines after a pause of a few seconds. The printing of lines continues until a "0" is sent to the sonde. The "0" will cause the sonde to immedaitly stop printing at a new line without echoing the "0". Which data values are printed can be set in the sonde menu Main > Report.