HIP:YSI-650MDS
Jump to navigation
Jump to search
- How to connect sonde and GPS to 650.
- To communicate with GPS:
- 1. In GPS menu, highlight SETUP -> INTERFACE
- 2. Set I/O FORMAT: NMEA OUT
- To communicate with GPS:
- How to calibrate sonde using 650 unit?
- (Example of pH)
- 1. Connecting 650 to sonde, highlight Sonde menu -> Calibrate -> ISE1 pH -> 1-point
- 2. Immerse the sonde in a buffer of known pH value and press Enter
- 3. Input the proper pH value for your calibration temperature, and press Enter
- How to harvest data?
- Run sonde from 650.
- 1. In 650 main menu, highlight Sonde run -> (under 650) start logging
- 2. Enter a file name and site description
- 3. 650 starts logging
- 4. When the study is complete, highlight the Stop logging selection and press Enter
- NOTE: You have to log the data to 650 memory, not to sonde memory, because GPS and barometer data cannot be logged to sonde memory.
- Run sonde from 650.
- How to update data on PC?
- 1. Connect 650 to stewie in HIP account.
- 2. Run EcoWatch, click on the sonde icon in the upper tool bar, and set the Comm port number to match your interface.
- 3. Highlight the Upload to PC selection in the 650 File menu and press Enter to view a list of the files. Highlight the file that you wish to transfer and press Enter.
- How to convert dat file to csv file?
- 1. Open dat file you uploaded on EcoWatch.
- 2. Set parameter of the dat file.
- Graphical example of setting parameter
- 3. Choose File->Export->CDF/WMF