*How to harvest data?
*How to connect sonde and GPS to 650.
**Connect sonde and GPS to 650.
[[Image:Connect_650_with_sonde_gps.jpg]]
[[Image:Connect_650_with_sonde_gps.jpg]]
**To communicate with GPS:
**To communicate with GPS:
***1. In GPS menu, highlight SETUP -> INTERFACE  
***1. In GPS menu, highlight SETUP -> INTERFACE  
***2. Set I/O FORMAT: NMEA OUT
***2. Set I/O FORMAT: NMEA OUT
*How to calibrate sonde using 650 unit?
**(Example of pH)
**1. Connecting 650 to sonde, highlight Sonde menu -> Calibrate -> ISE1 pH -> 1-point
**2. Immerse the sonde in a buffer of known pH value and press Enter
**3. Input the proper pH value for your calibration temperature, and press Enter
*How to harvest data?
**Run sonde from 650.
**Run sonde from 650.
***1. In 650 main menu, highlight Sonde run -> (under 650) start logging
***1. In 650 main menu, highlight Sonde run -> (under 650) start logging
Line 17: Line 24:
***2. Run EcoWatch, click on the sonde icon in the upper tool bar, and set the Comm port number to match your interface.
***2. Run EcoWatch, click on the sonde icon in the upper tool bar, and set the Comm port number to match your interface.
***3. Highlight the Upload to PC selection in the 650 File menu and press Enter to view a list of the files. Highlight the file that you wish to transfer and press Enter.
***3. Highlight the Upload to PC selection in the 650 File menu and press Enter to view a list of the files. Highlight the file that you wish to transfer and press Enter.
**How to convert dat file to csv file
**How to convert dat file to csv file
***1. Open dat file you uploaded on EcoWatch.
***1. Open dat file you uploaded on EcoWatch.
Line 26: Line 34:
*****[http://2.bp.blogspot.com/_3kd0zP9TXws/SLbTTNopQFI/AAAAAAAAAWU/fl37zXV8DeQ/s1600-h/Step+5.bmp Display data]
*****[http://2.bp.blogspot.com/_3kd0zP9TXws/SLbTTNopQFI/AAAAAAAAAWU/fl37zXV8DeQ/s1600-h/Step+5.bmp Display data]
***3. Choose File->Export->CDF/WMF
***3. Choose File->Export->CDF/WMF
**How to calibrate sonde using 650 unit?
***(Example of pH)
***Connecting 650 to sonde, highlight Sonde menu -> Calibrate -> ISE1 pH -> 1-point
***Immerse the sonde in a buffer of known pH value and press Enter
***Input the proper pH value for your calibration temperature, and press Enter.

Revision as of 17:52, 5 September 2008

  • How to connect sonde and GPS to 650.

    • To communicate with GPS:
      • 1. In GPS menu, highlight SETUP -> INTERFACE
      • 2. Set I/O FORMAT: NMEA OUT
  • How to calibrate sonde using 650 unit?
    • (Example of pH)
    • 1. Connecting 650 to sonde, highlight Sonde menu -> Calibrate -> ISE1 pH -> 1-point
    • 2. Immerse the sonde in a buffer of known pH value and press Enter
    • 3. Input the proper pH value for your calibration temperature, and press Enter
  • How to harvest data?
    • Run sonde from 650.
      • 1. In 650 main menu, highlight Sonde run -> (under 650) start logging
      • 2. Enter a file name and site description
      • 3. 650 starts logging
      • 4. When the study is complete, highlight the Stop logging selection and press Enter
      • NOTE: You have to log the data to 650 memory, not to sonde memory, because GPS and barometer data cannot be logged to sonde memory.
  • How to update data on PC?

      • 1. Connect 650 to stewie in HIP account.
      • 2. Run EcoWatch, click on the sonde icon in the upper tool bar, and set the Comm port number to match your interface.
      • 3. Highlight the Upload to PC selection in the 650 File menu and press Enter to view a list of the files. Highlight the file that you wish to transfer and press Enter.