Difference between revisions of "HIP:YSI-650MDS"

From Earlham CS Department
Jump to navigation Jump to search
Line 1: Line 1:
*How to harvest data?
+
*How to connect sonde and GPS to 650.
**Connect sonde and GPS to 650.
 
 
[[Image:Connect_650_with_sonde_gps.jpg]]
 
[[Image:Connect_650_with_sonde_gps.jpg]]
 
**To communicate with GPS:
 
**To communicate with GPS:
 
***1. In GPS menu, highlight SETUP -> INTERFACE  
 
***1. In GPS menu, highlight SETUP -> INTERFACE  
 
***2. Set I/O FORMAT: NMEA OUT
 
***2. Set I/O FORMAT: NMEA OUT
 +
 +
*How to calibrate sonde using 650 unit?
 +
**(Example of pH)
 +
**1. Connecting 650 to sonde, highlight Sonde menu -> Calibrate -> ISE1 pH -> 1-point
 +
**2. Immerse the sonde in a buffer of known pH value and press Enter
 +
**3. Input the proper pH value for your calibration temperature, and press Enter
 +
 +
*How to harvest data?
 
**Run sonde from 650.
 
**Run sonde from 650.
 
***1. In 650 main menu, highlight Sonde run -> (under 650) start logging
 
***1. In 650 main menu, highlight Sonde run -> (under 650) start logging
Line 17: Line 24:
 
***2. Run EcoWatch, click on the sonde icon in the upper tool bar, and set the Comm port number to match your interface.
 
***2. Run EcoWatch, click on the sonde icon in the upper tool bar, and set the Comm port number to match your interface.
 
***3. Highlight the Upload to PC selection in the 650 File menu and press Enter to view a list of the files. Highlight the file that you wish to transfer and press Enter.
 
***3. Highlight the Upload to PC selection in the 650 File menu and press Enter to view a list of the files. Highlight the file that you wish to transfer and press Enter.
 +
 
**How to convert dat file to csv file
 
**How to convert dat file to csv file
 
***1. Open dat file you uploaded on EcoWatch.
 
***1. Open dat file you uploaded on EcoWatch.
Line 26: Line 34:
 
*****[http://2.bp.blogspot.com/_3kd0zP9TXws/SLbTTNopQFI/AAAAAAAAAWU/fl37zXV8DeQ/s1600-h/Step+5.bmp Display data]
 
*****[http://2.bp.blogspot.com/_3kd0zP9TXws/SLbTTNopQFI/AAAAAAAAAWU/fl37zXV8DeQ/s1600-h/Step+5.bmp Display data]
 
***3. Choose File->Export->CDF/WMF
 
***3. Choose File->Export->CDF/WMF
**How to calibrate sonde using 650 unit?
 
***(Example of pH)
 
***Connecting 650 to sonde, highlight Sonde menu -> Calibrate -> ISE1 pH -> 1-point
 
***Immerse the sonde in a buffer of known pH value and press Enter
 
***Input the proper pH value for your calibration temperature, and press Enter.
 

Revision as of 12:52, 5 September 2008

  • How to connect sonde and GPS to 650.

Connect 650 with sonde gps.jpg

    • To communicate with GPS:
      • 1. In GPS menu, highlight SETUP -> INTERFACE
      • 2. Set I/O FORMAT: NMEA OUT
  • How to calibrate sonde using 650 unit?
    • (Example of pH)
    • 1. Connecting 650 to sonde, highlight Sonde menu -> Calibrate -> ISE1 pH -> 1-point
    • 2. Immerse the sonde in a buffer of known pH value and press Enter
    • 3. Input the proper pH value for your calibration temperature, and press Enter
  • How to harvest data?
    • Run sonde from 650.
      • 1. In 650 main menu, highlight Sonde run -> (under 650) start logging
      • 2. Enter a file name and site description
      • 3. 650 starts logging
      • 4. When the study is complete, highlight the Stop logging selection and press Enter
      • NOTE: You have to log the data to 650 memory, not to sonde memory, because GPS and barometer data cannot be logged to sonde memory.
  • How to update data on PC?

Upload data on PC.jpg

      • 1. Connect 650 to stewie in HIP account.
      • 2. Run EcoWatch, click on the sonde icon in the upper tool bar, and set the Comm port number to match your interface.
      • 3. Highlight the Upload to PC selection in the 650 File menu and press Enter to view a list of the files. Highlight the file that you wish to transfer and press Enter.